MLC Calibration

Semi-automatic MLC- und jaw calibration by means of Electronic Portal Imaging Device (EPID)

Preconditons:

  • Dosimetric calibrated EPID
  • Known central axis (CAx)

Electronic Portal Images (EPIs) are acquired from predefined field descriptions and the actual jaw and leaf positions are determined out of their particular penumbras. With this method, it is possible to verify and calibrate leaves and jaws within an accuracy of < 1 mm.

 

Fig.1: Abutted fidels collimated by similar jaws y1IEC an y2IEC    

Fig.2: Abutted fidels collimated by different jaws  x1IEC an y2IEC

Fig.3: Four quadrant  verification image of leaf and jaw junctions without collimator rotation after calibration

 

Key Personnel:

  • Mag. Deutschmann H., Institute for Research and Development on Advanced RadiationTechnologies (radART), Paracelsus Medical University, Salzburg, Austria
  • DI Mitterlechner B., Institute for Research and Development on Advanced RadiationTechnologies (radART), Paracelsus Medical University, Salzburg, Austria

 

Posters:

  • Dosimetric exact multileaf collimator calibration by means of aSi:H Electronic Portal Imaging Device, Mitterlechner B.2, Sedlmayer F.1,2, Deutschmann H.1,2 (IGRT Salzburg 2011, Tagung der Österreichischen Gesellschaft für Medizinische Physik – ÖGMP, Salzburg 2010)

 Oral Presentations:

  • Semi-automatische MLC- und Blendenkalibrierung mittels Electronic Portal Imaging Device, Mitterlechner B.2, (Tagung der Österreichischen Gesellschaft für Medizinische Physik – ÖGMP, Salzburg 2010)

 

1 Institute for Research and Development on Advanced RadiationTechnologies (radART), Paracelsus Medical University, Salzburg, Austria

2 Department of Radiotherapy and Radio-Oncology at the Salzburger Landeskliniken, Salzburg, Austria

 